Centrifuge offers protection in health care environments with a sealed, transparent polycarbonate cover for aerosol and splash protection.
Designed for physician's offices and clinical labs for blood, fecal, and high density sedimentation separation.
Assure product protection and reliability by having repairs and maintenance performed by authorized service centers.
Place unit on a stable surface, plug into a grounded receptacle, and remove protective wrap from the cover.
Operate from approved AC sources, use grounded receptacles, and never remove the grounding prong.
Depress detente fasteners to open the cover; ensure both sides lock when closing for operation.
Load 10 ml or 15 ml tubes in a balanced manner, using adapter cushions if needed.
Use the on/off switch for basic operation or the 60-minute timer for timed cycles.
Always operate with cover locked, balanced load, and never open during operation or before rotor stops.
Frequently clean all parts and always replace the rotor collar gasket if worn or damaged.
Clean bowl area with disinfectant; clean rotor with hypochlorite solution, rinse, and dry. Do not autoclave rotor.
Check rotor periodically for cracks; replace if necessary with the specified V7000 rotor.
Always unplug power before servicing and wear appropriate protective gear and disinfectant.
SST tubes contain a gel that separates clot from serum for chemistry testing.
Procedure for processing SST tubes: collect, mix, allow to clot, spin, and transport.
List of available replacement parts including rotors, shields, gaskets, and adapters.