Rinse the pH electrode with some pH 4.01
or 10.01 buffer solution (2
nd
calibration
point).
Dip the electrode into the pH 4.01 (or
pH 10.01) buffer solution.
Stir gently and wait until the stability
indicator "" stops flashing.
Press CFM to confirm the acceptance of
the second buffer solution.
pH calibration is now complete and the meter
returns to the measurement mode.
PROCEDURE FOR HI 931000
Turn the meter on.
Note: the stability indicator "" blinks when
the electrode is not immersed in the
solution.
Press CAL and the display will flash "7.01 ".
Remove the protective cap from the elec-
trode, rinse the tip of the electrode with
some pH 7.01 solution, then immerse it
into a pH 7.01 buffer solution; stir gen-
tly and wait a couple of minutes for
thermal equilibrium.
Note: the electrode should be submerged
approximately 4 cm (1½") into the
solution.
When electrode has stabilized, the sta-
bility indicator "" stops flashing.
Press CFM to confirm the first buffer solution.
ins
e
rum
n
s
ON
O
FF
C
A
L
CFM
H
I
9
2
1
4
s
tic
k
p
H
m
e
te
r
HI 7004
ON
OFF
CAL
CFM
HI 7007
4 cm
(1
½
")
CFM
PROCEDURE FOR HI 9214:
Turn the meter on.
Note: the stability indicator "" blinks when
the electrode is not immersed in the
solution.
Remove the protective cap from the pH/°C
electrode, rinse the tip of the electrode
with some pH 7.01 solution, then immerse
it into a pH 7.01 buffer solution; stir
gently and wait a couple of minutes for
the reading to stabilize.
Note: the electrode should be submerged
approximately 4 cm (1½") into the
solution.
Press CAL. The display will flash "7.01 ".
Wait until the stability indicator "" stops
flashing.
Press CFM to confirm the first buffer solu-
tion.
If the ON/OFF key is pressed at this point, the calibration process
will end with only the offset (one-point) calibration. For best
accuracy, however, it is recommended that two-point calibration
is performed.
The LCD will now flash "4.01 ". If the
CAL key is pressed, the slope calibration
will change to "10.01 ".
ON
OFF
ins
e
rum
n
s
O
N
O
FF
C
A
L
CFM
H
I
9
2
1
4
s
tic
k
p
H
m
e
te
r
HI 7007
4 cm
(1
½
")
ins
e
rum
n
s
ON
OFF
C
A
L
C
F
M
H
I
9
2
1
4
s
t
ic
k
p
H
m
e
t
e
r
CAL
CFM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 8424 and is the answer not in the manual?

Hanna Instruments HI 8424 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 8424
CategoryDigital multimeter
LanguageEnglish

Related product manuals