pH CALIBRATION SOLUTIONS
HI 70004P pH 4.01 Buffer Sachets, 25 x 20 mL
HI 7004M pH 4.01 Buffer Solution, 230 mL
HI 7004L pH 4.01 Buffer Solution, 460 mL
HI 7006M pH 6.86 Buffer Solution, 230 mL
HI 7006L pH 6.86 Buffer Solution, 460 mL
HI 70007P pH 7.01 Buffer Sachets, 25 x 20 mL
HI 7007M pH 7.01 Buffer Solution, 230 mL
HI 7007L pH 7.01 Buffer Solution, 460 mL
HI 7009M pH 9.18 Buffer Solution, 230 mL
HI 7009L pH 9.18 Buffer Solution, 460 mL
HI 70010P pH 10.01 Buffer Sachets, 25 x 20 mL
HI 7010M pH 10.01 Buffer Solution, 230 mL
HI 7010L pH 10.01 Buffer Solution, 460 mL
pH CALIBRATION SOLUTIONS IN FDA APPROVED
BOTTLE
HI 8004L pH 4.01 Buffer Solution, 460 mL
HI 8006L pH 6.86 Buffer Solution, 460 mL
HI 8007L pH 7.01 Buffer Solution, 460 mL
HI 8009L pH 9.18 Buffer Solution, 460 mL
HI 8010L pH 10.01 Buffer Solution, 460 mL
ELECTRODE STORAGE SOLUTIONS
HI 70300M Storage Solution, 230 mL
HI 70300L Storage Solution, 460 mL
ELECTRODE STORAGE SOLUTIONS IN FDA
APPROVED BOTTLE
HI 80300M Storage Solution, 230 mL
HI 80300L Storage Solution, 460 mL
ELECTRODE CLEANING SOLUTIONS
HI 70000P Electrode Rinsing Sachets, 25 x 20 mL
HI 7061M General Cleaning Solution, 230 mL
HI 7061L General Cleaning Solution, 460 mL
HI 7073M Protein Cleaning Solution, 230 mL
HI 7073L Protein Cleaning Solution, 460 mL
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDE
HI 8014, HI 8314, HI 8424, HI 8915, HI 9214 AND
HI 9622
°C Symbol indicates the meter is in temperature mode. A flashing
°C symbol in HI 9622 means that the temperature probe is not
connected or the cable is damaged
pH Symbol indicates the meter is in pH mode.
HI 8014, HI 8314, HI 8424, HI 8915 AND HI 9622
mV Symbol indicates the meter is in mV mode.
HI 8424
E1 Out of range in pH scale
E2 Out of range in °C scale
E3 Out of range in mV scale
E4 Erroneous buffer solution used for offset calibration or out of order
electrode
E5 Erroneous buffer solution used for slope calibration or out of order
electrode
HI 8915
S Symbol to indicate the display is frozen (Hold function).
HI 9214 AND HI 931000
∆∆
∆∆
∆ Stability indicator
Ec Wrong buffer: Change the buffer solution;
Unfit electrode: Clean and condition the electrode.
Eb Low battery: Replace the battery.
Er Over-Range: Dry electrode, repeat conditioning.
HI 9622
pH---- Out of range in pH scale
mV---- Out of range in mV scale
Er-7 Erroneous pH 7 buffer solution used for offset calibration or
out of order electrode
Er-4 Erroneous pH 4 buffer solution used for slope calibration or
out of order electrode
Er-10 Erroneous pH 10 buffer solution used for slope calibration or
out of order electrode
BAT Low battery