Hanna Instruments HI5221 - Sample ID Configuration

Hanna Instruments HI5221
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
28
Press
while in pH Measure mode.
Press
.
• Use
or to select the Calibration
option.
Press
and use or to
highlight the Clear Calibration option.
• Press
to clear calibration. A pop-up menu
will be displayed asking for confirmation (when a
calibration is available).
• Press
to confirm or press to escape
without saving and return to the Calibration options.
Sample ID
This option allows the user to assign an identification
number/name. Two Sample ID options are available:
ID Increment and Edit Sample ID.
ID Increment
Two choices are available for the sample ID:
None - the sample ID will be fixed and it can be set
alphanumeric (see Edit Sample ID).
Automatic - the sample ID will automatically increment
by one for each new log lot.
To set the ID Increment mode:
Press
while in pH Measure mode.
Press
.
• Use
or to select the Sample ID option.
• Press
and use or to highlight the ID Increment option.
• Press
/ as desired.
• Press
to return to previous menu.
pH SETUP
To clear Calibration: