EasyManuals Logo

Hanna Instruments HI774 User Manual

Hanna Instruments HI774
4 pages
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #4 background imageLoading...
Page #4 background image
Tips for an Accurate Measurement
Ensure the sample does not contain any debris.
Whenever the cuvette is placed into the checker, it must be
dry outside and free of fingerprints, oil and dirt.
Wipe the cuvette thoroughly with HI731318 microfiber
cleaning cloth or a lint-free cloth prior to insertion.
Shaking the cuvette can generate bubbles, causing higher
readings. To obtain accurate measurements, remove such
bubbles by swirling or by gently tapping the cuvette.
Do not let the reacted sample stand too long after reagent
has been added, as accuracy will be affected.
Discard the sample immediately after the reading has been
taken or the glass might become permanently stained.
Battery Replacement
To save the battery, the checker shuts down after 7 minutes of non-use and
2 minutes after reading.
A fresh battery lasts for a minimum of 5000 measurements. When the battery is
drained, the instrument displays “bAd” then ”bAt”, and turns off.
To replace the battery, follow the next steps:
1.
Press and hold the ON/OFF button to turn the checker off.
2.
Turn the instrument upside down and use a screwdriver to unfasten the screw and
remove the battery cover.
3. Remove the old battery, replace it with a new 1.5V AAA battery, inserting the
negative end first.
4.
Replace the battery cover, fasten and tighten the screw.
Accessories
Low Range
w ww . . co m
information@itm.com1.800.561.8187

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI774 and is the answer not in the manual?

Hanna Instruments HI774 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI774
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals