Do you have a question about the Hart Scientific 9100 HDRC and is the answer not in the manual?
Overview of the Hart Scientific Model 9100 HDRC Dry-well Calibrator, its purpose, and features.
Details the operating range, accuracy, stability, heating/cooling times, probe hole sizes, depth, power, size, and weight of the calibrator.
Specifies the recommended operating temperature, humidity, pressure, voltage, vibration, and altitude conditions for safe use.
Outlines the product warranty, limitations, and contact information for service or assistance.
Provides critical safety precautions for operating the dry-well calibrator to prevent injury and equipment damage.
Instructions for carefully unpacking the dry-well and verifying all components are present.
Guidance on placing the calibrator, connecting the power cord, and turning it on for initial operation.
Details on connecting the power cord to the appropriate voltage and turning the unit on.
Description of the rear panel components including power cord, power switch, and fan.
Explanation of the prop's location, function, and usage for inclined positioning.
Details the front panel components: well block, controller display, and controller buttons.
Instructions for placing the calibrator, connecting power, and initial power-on sequence.
Procedure for setting the desired temperature using the controller buttons and understanding temperature stabilization.
Overview of controller programmability, software lock, and accessing the menu levels.
Steps to unlock the controller parameters for modification and relock them for security.
Method to change the displayed temperature units between Celsius (°C) and Fahrenheit (°F).
Describes different methods for calibrating test probes, including direct and comparison calibration.
Procedure for calibrating a probe by directly comparing its reading to the dry-well's displayed temperature.
Method for calibrating a probe against a known reference probe for improved accuracy.
Guidance on calibrating several probes simultaneously, noting stabilization time and stem effect.
Information on understanding the thermal properties of the dry-well for optimal calibration.
Explanation of vertical temperature gradients within the test well and how to minimize them.
Details on the time required for the dry-well to stabilize and achieve accuracy at set temperatures.
Steps to ensure all controller parameters are set correctly before proceeding with calibration.
Accuracy specifications and the procedure for calibrating the dry-well.
Detailed steps to calculate and set the new span parameter for calibration.
Instructions for calculating and setting the zero parameter for accurate calibration.
Procedure for checking the accuracy of the dry-well at various set-points and repeating adjustments if needed.
Guidelines for routine maintenance, cleaning, handling, and general care of the dry-well calibrator.
Common problems encountered during operation and suggested solutions for fixing them.
Information regarding compliance with European directives (EMC and Low Voltage) for safety and electromagnetic compatibility.
Details on testing for electromagnetic compatibility according to the EMC Directive.
Information on compliance with the Low Voltage Directive and relevant IEC safety standards.
Schematic diagram illustrating the electrical connections and components of the 9100 Dry-Well.
| Model | 9100 HDRC |
|---|---|
| Type | Dry-Well Calibrator |
| Communication Interface | RS-232 |
| Display | °C or °F |
| Power Supply | 115 VAC or 230 VAC |