by the legislator, such as, in
Germany, the “Medizinprodukte-
Betreiberverordnung” (Medical
Device Operating Regulation).
15.2 Instructions for the
calibration check
The calibration check can be carried
out by the competent authorities or
authorised maintenance providers
against compensation. A device
function check can be carried out
on people or using a suitable simu-
lator. Calibration check involves
testing for leak tightness of the
pressure system and possible devia-
tions of the pressure reading.
Remove at least one battery in
order to switch to calibration mode.
Hold down on the START / STOP
button and then insert the battery.
Release the button and, after a few
moments, two zeros will appear
one above each other on the dis-
play. Instructions on the calibration
check will be supplied on request
to competent authorities or
authorised maintenance providers
by HARTMANN.