HARVEY CHEMICLAVE 5000
OWNER AND OPERATOR GUIDE
OPERATING INSTRUCTIONS
FILLING AND CHECKING SOLUTIONS
The filling opening is located on top of the Chemiclave (see figure 1 below). The dust cover
should be removed only for filling or checking solution level.
With the control knob in the "Depressurize" position, fill the solution reservoir with approximately
500 ml of Vapo Steril. The level of solution may be determined visually. A measuring cup is
included for your convenience.
Since each cycle of the Chemiclave uses 15.7 ml of Vapo-Steril Solution, refilling of the solution
reservoir is required after approximately 30 cycles. The "Solution: light (see figure 1 below)
signals when the solution reservoir needs refilling.
When the solution reservoir is empty, the system will not build pressure and refilling with Vapo-
Steril is necessary. The solution reservoir may be refilled at any time, whether the Chemiclave is
hot or cold. If the solution reservoir is empty, turn the control knob to "Depressurize" before
refilling (see figure 1 below)
Before refilling solution reservoir, dispose of exhausted solution by removing condensate tray
(see Figure 1) and discarding contents. The condensate is contaminated and chemically altered
and must not be reused. IF REUSED, CONDENSATE WILL NOT STERILIZE AND MAY
DAMAGE YOUR CHEMICLAVE.
ACTIVATING CHEMICLAVE
Connect the Chemiclave to 115 Volts, 60hz only. Turn on the power switch (see figure 1 below).
Two lights (see figure 1 below) indicate your Chemiclave is functioning properly. The "Power"
light will remain on until turned off at the end of the day.
The "Temperature" light will remain on up to 15 minutes during the initial heating period. Your
Chemiclave is designed to operate all day with minimal power consumption (only 500 watts, as
against 1000 to 1250 watts for most other units). To save valuable time, leave it on to avoid
"preheating" for each load.
When the "Temperature" light goes out initially, the Chemiclave is ready for use. After initial
start-up, the "Temperature" light will illuminate intermittently, as the thermostat automatically
maintains the optimum temperature for sterilization (132 degrees c). NO MANUAL
ADJUSTMENTS ARE NECESSARY TO MAINTAIN THIS TEMPERATURE.
Question and Answer IconNeed help?

Do you have a question about the HARVEY CHEMICLAVE 5000 and is the answer not in the manual?

HARVEY CHEMICLAVE 5000 Specifications

General IconGeneral
BrandHARVEY
ModelCHEMICLAVE 5000
CategorySteriliser
LanguageEnglish

Summary

Filling and Checking Solutions

Solution Reservoir Refilling

Instructions for filling the solution reservoir with Vapo-Steril and refilling signals.

Condensate Disposal

Procedure for disposing of exhausted condensate and warnings against reuse.

Activating the Chemiclave

Power On and Indicator Lights

Connecting to power, identifying Power and Temperature lights, and initial heating.

Operation and Temperature Maintenance

How the Chemiclave operates at optimal sterilization temperature (132°C) automatically.

Preparing Items for Sterilization

Instrument Cleaning and Rinsing

Guidelines for cleaning, rinsing, and drying instruments to prevent contamination and residue.

Loading Instruments for Sterilization

Instructions for placing small objects, handpieces, and items in bags for sterilization.

Sterilizing Process

Cycle Setup and Operation

Instructions for loading the tray, setting the control knob, and the automatic 20-minute cycle.

Sterilization Phases Explained

Explanation of the condensation and vaporization phases during the sterilization cycle.

Completing the Cycle

Cycle Completion Signals

How to recognize cycle completion via buzzer and 'Sterile' light and depressurize.

Safe Door Opening and Unloading

Ensuring chamber exhaustion before opening the door and removing contents immediately.

Helpful Suggestions and Parts

General Usage Suggestions

Tips for using the Chemiclave in professional settings for optimal sterilization.

Door Gasket Care

Recommendations for replacing and maintaining the door gasket for proper sealing.

Maintenance and Cleaning

Exterior Surface Cleaning

Daily cleaning of exterior metal surfaces with Harvey Metal Polish.

Chamber and Tray Cleaning

Periodic cleaning of the chamber and tray to remove debris from Vapo-Steril.