Hold T + H together for 3 seconds to enter menu
- Use T/H buttons to choose parameters.
- Use U/D buttons to change set values.
- Hold T + H together for 3 seconds to save changes.
hhi: high humidity alarm
Started: set value +10
Delay time: 20 minutes
Notification: Alr 5
Alarm: continuous
hlo: low humidity alarm
Started: set value -10
Delay time: 20 minutes
Notification: Alr 6
Alarm: intermittent
(f)thi: high temperature alarm
Started: set value +1.5
Delay time: 2 minutes
Notification: Alr 1/3
Alarm: continuous
(f)tlo: low temperature alarm
Started: set value -4.0
Delay time: 30 minutes
Notification: Alr 2/4
Alarm: intermittent
tco: temperature calibration value
CAUTION! tco is a critical parameter. Changing this can
results in chick loss. Use qualified and calibrated glass
or electronic thermometers for calibrating.
Check temperature values regularly.
hco: humidity calibration value
CAUTION! hco is a critical parameter. Changing this can
results in chick loss. Use qualified and calibrated glass or
electronic thermometers for calibrating.
Check humidity values regularly.
US2: USB connected device
no: no connected device
th: second sensor connected (room/calibration)
ıot: iot module connected
Notification: Alr 15 (usb connection error)
Below parameters active if Us2 is selected “th”
(f)rtl: second sensor low temperature alarm
Started: below 65°F (18°C)
Delay time: no
Notification: Alr 10
Alarm: intermittent
(f)rth: second sensor high temperature alarm
Started: over 83°F (28°C)
Delay time: no
Notification: Alr 9
Alarm: intermittent
rhl: second sensor low humidity alarm
Started: below 10 RH
Delay time: no
Notification: Alr 14
Alarm: intermittent
rhh: second sensor high humidity alarm
Started: over 85 RH
Delay time: no
Notification: Alr 13
Alarm: intermittent
rAt: second sensor temperature alarm activation
On: alarms activated
Off: alarms deactivated
rAh: second sensor humidity alarm activation
On: alarms activated
Off: alarms deactivated