HeartSine samaritan PAD - User Manual

HeartSine samaritan PAD
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
SAM 350P
Semi-automatic defibrillator
EN-US
Important:
Please watch online training videos
heartsine.com/videos
HeartSine
®
samaritan
®
PAD
user manual

Other manuals for HeartSine samaritan PAD

Questions and Answers

  • R
    Ronnie CoxJul 29, 2025
    What to do if HeartSine Medical Equipment has a red status indicator and beeping while device is on?
    • S
      Scott DukeJul 29, 2025
      If the HeartSine Medical Equipment shows a red status indicator and beeps while on, you should immediately replace the Pad-Pak or find another defibrillator. If you don't have a spare Pad-Pak or alternative defibrillator, the device will keep analyzing the patient's heart rhythm and tell you when to do CPR, but it won't be able to deliver a shock.
  • W
    William NolanAug 3, 2025
    What to do if HeartSine samaritan PAD shows a 'Low battery' warning?
    • M
      Manuel AdamsAug 3, 2025
      If your HeartSine Medical Equipment displays a “Low battery” warning, replace the battery as soon as you can. The AED will still work, but it might have fewer than 10 shocks left. Prepare a spare Pad-Pak for quick use and order a new Pad-Pak immediately.
  • K
    Kevin LewisAug 7, 2025
    How to fix flashing red status indicator or continual beeping on HeartSine Medical Equipment?
    • J
      Jessica BlackburnAug 7, 2025
      First, check the expiration date on your Pad-Pak. If it's expired, replace it right away. If not, turn the HeartSine samaritan PAD on and listen for the voice prompt “Call for medical assistance”. Then, turn the device off again. If the problem persists, contact your Authorised Distributor or HeartSine Technologies immediately.
  • D
    Debra HarrellAug 12, 2025
    What to do if HeartSine samaritan PAD says 'Check pads'?
    • W
      Wanda WolfeAug 13, 2025
      If the HeartSine Medical Equipment displays a “Check pads” prompt, make sure the pads are fully attached to the patient as shown in the diagram, and that the skin is clean, dry, and hair-free. Readjust the pads if needed. If the message continues, remove and reinsert the Pad-Pak. If the message still appears, use another defibrillator and continue CPR.
  • C
    Christopher EvansAug 18, 2025
    Why does my HeartSine samaritan PAD beep three times when turned off?
    • H
      hwashingtonAug 18, 2025
      The HeartSine Medical Equipment beeps rapidly three times when turned off or after a weekly self-test because it's outside the specified operating temperature. Return your AED to a temperature between 0°C and 50°C (32°F to 122°F). Then, check if the beeping stops.
  • R
    rwillisAug 23, 2025
    What to do if HeartSine samaritan PAD displays a 'Device service required' warning?
    • M
      Michelle WalshAug 23, 2025
      If the HeartSine Medical Equipment shows a “Device service required” warning, immediately find an alternative defibrillator. Do not try to fix the device yourself, as it cannot be modified. Contact HeartSine Technologies or your authorized distributor right away.

HeartSine samaritan PAD Specifications

General IconGeneral
TypeAutomated External Defibrillator (AED)
IP RatingIP56
Battery Life4 years (standby)
Energy150 J (biphasic)
Operating Temperature0°C to 50°C (32°F to 122°F)
WaveformBiphasic
Energy SelectionAutomatic
Charge TimeLess than 8 seconds
Voice PromptsYes
CPR CoachingYes
Battery TypeLithium Manganese Dioxide (LiMnO2), non-rechargeable
Data RecordingYes
Relative Humidity5% to 95% non-condensing
AltitudeUp to 4, 572 m (15, 000 ft)
Pediatric UseYes