Montageanleitung
Mounting Instructions
LB 302
LB 382
mehrteilig
Multi
-
Section
7
/20
1
4
Question and Answer IconNeed help?

Do you have a question about the HEIDENHAIN LB 302 and is the answer not in the manual?

HEIDENHAIN LB 302 Specifications

General IconGeneral
ProtectionIP 53
Operating Current≤ 200 mA
Protection ClassIP53
Signal Period20 µm
Power Supply5 V ±5%
Operating Temperature0 °C to +50 °C
Operating Voltage5 VDC ±5%
Storage Temperature-20°C to 70°C

Summary

General Safety Warnings

Mounting and Commissioning by Qualified Personnel

Installation and commissioning must be performed by a qualified specialist following safety regulations.

Electrical Safety Precautions

Ensure the system is disconnected from power before connecting or disconnecting. Use only when voltage-free.

Product Components

Main Components A, B, C, D

Illustrates the main components of the system, including the scanning unit and housing sections.

Components Available Separately

Identifies items that are not included and must be ordered separately, like adapter cables.

Component Details

Shipping Brace (T)

Illustrates the shipping brace and its use for securing the scanning unit during transport.

Connecting Cable and Mounting Gauge

Details the connecting cable and the mounting gauge used for installation and alignment.

Included Parts and Kits

Assembly Instructions and Small Parts

Details the mounting instructions booklet (A) and small parts kit (K) included in the delivery.

Scale Tape and Sealing Lips

Identifies the scale tape (M), sealing lips (L), and bearing strips (B) supplied with the product.

End Sections and Seal

Shows the included end sections (E1, E2) and seal (D) for the scale housing.

Small Parts Details

Tools and Fasteners

Lists and describes small parts such as the scale tape puller (Ew), sealing lip clamp (O), and plugs (PF).

Reference Mark Components

Details components for reference marks, including the slider (S) and label (RM).

Mounting Procedure and Guidelines

Mounting Position Requirements

Select mounting location to prevent scanning unit contact and protect sealing lips from contamination.

Vertical Mounting Considerations

For vertical installation without compressed air, remove the drain screw.

Mounting Precautions

Preventing Contamination

Ensure no contamination enters the scale housing during the mounting process.

Scale Tape Handling

Do not touch the graduation side of the scale tape and avoid bending it.

Cable Outlet Modification

Electrostatic Discharge (ESD) Protection

Protect electronics from static charge using a grounded bracelet during handling.

Printed Circuit Board (PCB) Adjustment

Tilt, pull out, and rotate the PCB by 180°, inserting the connector side first. Avoid pinching wires.

Reference Mark Positioning

Selecting Reference Mark Locations

Reference marks can be selected at mounting holes or at intervals of n x 50 mm.

Installing Reference Mark Slider

Insert the reference mark selector plate and slide the red reference mark slider to the desired position.

Standard Version Dimensions

Dimensional Drawings and Tolerances

Provides detailed drawings with dimensions, tolerances (ISO 8015, ISO 2768-mH), and component labels.

Mirror-Imaged Version Dimensions

Dimensional Drawings and Tolerances (Mirror-Imaged)

Presents detailed drawings with dimensions and tolerances for the mirror-imaged configuration.

Mounting Tolerances

Mounting Possibilities and Tolerances

Details mounting possibilities and specific tolerances for positioning and alignment.

Horizontal Mounting Attitude

Confirms that horizontal mounting is a permissible installation method.

Mounting Surface and Hole Details

Mating Surface and Mounting Holes

Specifies requirements for mating surfaces and details of threaded mounting holes and screws.

Mounting Surface Preparation

Emphasizes that mounting surfaces must be clean and free of paint for proper installation.

Mounting Housing Sections

Initial Housing Section Preparation

Remove end cap, scale clamp, and protecting cap. Align end section to machine guideway and screw down.

Joining Multiple Housing Sections

Use a washer for recessed elements. Slide seal with shipping protection off. Drill holes if needed.

Assembling Housing Sections

Joining Subsequent Housing Sections

Place next section, align, mark holes, maintain gap 'b' using spacer Z, insert seal D.

Securing Aligned Housing

Align the attached housing to the machine guideway F and screw down. Continue with remaining sections.

Finalizing Housing Section Attachment

Securing the End Cap and Cover Plate

Remove end cap and cover plate Q. Loosen screw Ss. Remove tensioning device and slide off stop plate KA.

Attaching the Final End Section

Attach the end section to the housing section, align, and screw down. Follow same procedure for other sections.

Inserting Bearing Strips

Using the Scale Tape Puller

Use the scale tape puller Ew to insert bearing strips. Hook strips with beveled corners pointing upwards.

Proper Seating of Bearing Strips

Slide the bearing strip into the housing, ensuring it is properly seated in the groove.

Scale Tape Mounting Guidelines

Scale Tape Handling and Cleaning

Do not touch the graduation; clean contaminated surfaces with lint-free cloth and spirit or alcohol.

Recording Scale Tape Information

Write down the shortening factor Vk and scale-tape serial number from the aluminum label.

Orientation of Scale Tape

Ensure the square hole aligns with end section E2. Observe correct orientation of the graduation side.

Scale Tape Installation and Securing

Inserting the Scale Tape

Insert scale tape puller Ew, hook the tape, slide it into the housing, ensuring correct groove seating.

Securing the Scale Tape with Tensioning Device

Screw out tensioning screw SM, hook tape, slide in with device, and tighten screws (1 Nm).

Final Scale Tape Securing

Initial Scale Tape Affixing

Unscrew clamping screws Sk, secure holder KM with screws ST. Loosely affix scale tape with clamping screws Sk.

Tensioning the Scale Tape

Loosen ST, tighten Sk for no play (0.1 Nm). Tighten ST (1 Nm). Tighten SM until no play, then tension by 10 revolutions.

Scale Tape Tension Adjustment

Final Clamping Screw Tightening

Tighten clamping screws SK to approximately 1.5 Nm, ensuring the 2 Nm torque limit is not exceeded.

Releasing Scale Tape Tension

Loosen the scale tape by 11 revolutions of the tensioning screw SM.

Inserting Sealing Lips

Preparing and Lubricating Sealing Lips

Sealing lips are preloaded. Mark length (L x 0.99). Lubricate the inside with sealing lip grease FT.

Sealing Lips Insertion Details

Initial Sealing Lip Placement

Pull in both sealing lips, letting them protrude 1 mm. Loosen with fingers if stiff.

Orienting Sealing Lips

Use reference mark slider S to orient sealing lips outward over ~500 mm. Ensure correct relative positioning.

Sealing Lips Trimming

Positioning and Cutting Sealing Lips

At the second end, pull lips out 1 mm past marking. Orient outward ~500 mm. Cut lips at the marking.

Final Sealing Lip Alignment

Final Sealing Lip Orientation

Orient sealing lips outward over the entire length, ensuring 1 mm protrusion. Do not move reference mark selector plate R.

Securing Sealing Lips (End Section E2)

Attaching Stop Plate and End Cap

Slide stop plate KA onto end section. Secure end cap with 3 screws (1 Nm). Ensure lips are seated correctly.

Using Sealing Lip Clamp

Loosen screw SO on end cap K2. Insert sealing lip clamp O under the end cap. Tighten screw SO flush.

Installing the Scanning Unit

Securing the Scanning Unit

Use the included shipping brace to secure the scanning unit during mounting and moving to prevent damage.

Initial Brace Engagement

Click-in the shipping braces. Adjust screw S if necessary for loosening or tightening.

Scanning Unit Insertion and Mounting

Sliding Unit into Housing

Carefully insert unit with first brace, slide into housing until second brace can be clicked in.

Final Unit Placement and Securing

Connect braces, slide unit to mounting surface, lightly tighten screws. Remove shipping brace.

Scanning Unit Adjustment and Connection

Setting the Scanning Gap

Set the scanning gap using the mounting gauge (ID 772141-01). Maintain gap over the entire measuring length.

Connecting Adapter Cable and Final Checks

Connect the adapter cable. Tighten screws evenly. Check mounting tolerances over the entire measuring length.

Securing Sealing Lips (End Section E1)

Preparing End Cap and Clamp

Screw on the end cap and loosen clamping screw So. Insert sealing-lips clamp O under the housing end cap.

Tightening Sealing Lip Clamp Screw

Tighten screw So so that it is flush.

Final Installation Steps

Shielding Resistance Check

Check shielding by measuring resistance between connector housing and scale housing. Desired value: 1 Ω max.

Connecting to Display Unit

Connect the encoder to a HEIDENHAIN display unit and check function over the entire traverse range.

Attaching ID Label

Attach the product ID label (T2) to the scale.

Scale Tape Tensioning Procedure

Initial Tensioning and Display Reset

Slide unit to end, reset display to zero. Remove plug PF. Tighten tensioning screw SM for pretension (approx. 50 µm).

Calculating Tension Value

Loosen SM until display stops changing. Reset display to zero. Calculate tension value V using Vk and distance Xv.

Finalizing Scale Tape Tension

Adjusting Tape Tension

Increase tape tension until display shows calculated value. Tighten clamping screws SS (2 Nm).

Replacing Plug and Cover

Replace plug PF. Attach cover Q with screws M2.5 x 5 (0.6 Nm).

Linear Error Compensation

Applying Linear Compensation

Apply linear error compensation up to ± 100 µm/m using the tape tensioning device (see pages 32 and 33).

Machine Measurement Setup

Set up a comparator system, like a laser interferometer, in the workpiece plane to measure the machine tool.

Calculating Compensation Value

Preparing for Calculation

Remove cover Q and plug PF. Slacken screw SS. Prepare for compensation value calculation.

Determining Compensation K

Calculate compensation value K by measuring distance XK and multiplying with linear compensation value LK.

Protective Measures and Air Connection

Protection Against Contamination

Fit a protective cover with a seal between the encoder and mounting surface if contamination risk is high.

Compressed Air Requirements

Use clean, dry compressed air (1 bar) only via the connecting piece. Connect at scanning unit or end sections.

Mechanical Specifications

Measuring Standard and Reference Marks

Details the AURODUR graduation, grating period, thermal expansion, and reference mark selection.

Performance and Environmental Data

Covers traversing speed, acceleration, protection type (IP 53/64), operating/storage temperatures.

Connecting Cable Bending Radii

Specifies permissible bending radii for the connecting cable under frequent flexing and rigid configuration.

Electrical Connection (LB 302/LB 302 C)

9-Pin HEIDENHAIN Connector Pinout

Details the pin assignments for the 9-pin HEIDENHAIN connector, including signal lines and wire colors.

Signal Path and Noise Sources

Illustrates the signal path and identifies potential noise sources in the system.

Electrical Data (LB 302/LB 302 C)

Power Supply and Output Signals

Specifies DC 5 V power supply and details output signal characteristics (sine waves, phase).

Incremental and Reference Signals

Provides signal amplitudes for incremental signals (I1, I2) and reference mark signal (I0) at 1 kΩ load.

Electrical Connection (LB 382/LB 382 C)

12-Pin HEIDENHAIN Connector Pinout

Details the pin assignments for the 12-pin HEIDENHAIN connector and coupling, including signal lines and colors.

Sensor Line and Shielding

Explains that the sensor line is internally connected to the supply line and the shield is on the housing.

Electrical Data (LB 382/LB 382 C)

Power Supply and Output Signals

Specifies DC 5 V power supply and output signal characteristics (sinusoidal A, B, R) with terminating resistor.

Incremental and Reference Signal Amplitudes

Provides signal amplitudes for incremental signals (A, B) and reference mark signal (R) with terminating resistor.