Non Contact Infrared Body
Thermometer
Instructions for use
9
Forehead mode: press On/Scan button to open the device, the thermometer is
aligned with the middle of the forehead to measure body temperature (between the
eyebrows above) and keep the vertical distance, press the On/Scan button, the
position lamp indicates measurement site with light spot on the forehead,
temperature display immediately, see figure 3.
Forehead Scan mode: gently position the probe flush (flat) on the center of the
forehead, midway between the eyebrow and the hairline. Press and hold the
On/Scan button. Lightly slide the thermometer across the fore-head keeping the
sensor flat and in contact with the skin until you reach the hairline, release the
On/Scan button and remove the thermometer from the head, then the temperature
will display on the screen. The whole process takes 3~10 seconds.
When the ambient temperature of the thermometer changes too much, such as
moving the Thermometer from one place of lower temperature to another place of
higher temperature, Allow the thermometer to remain in a room for 30 minutes where
the temperature is between 5°C to 40°C.
The ambient temperature around the test person should be stable, should keep away
from the larger flow fan, air-conditioning vents and so on.
When people moving from one place of lower temperature to another place of
higher temperature, should at least remain in the test environment more than 5
minutes, to be consistent with the ambient temperature after the re-measurement.
Wait at least 1 second for the next measurement. If the continuous measurement of
five times, it is recommended to wait at least 30 seconds and then continue
measurement.
You cannot use the thermometer in place where the sun is strong.
Figure 3-Measuring position and distance