3.3 Integrating the Pulse Output from a Power Meter
37
1
2
3
4
5
6
7
8
9
10
5
6. Connect the 9641 CONNECTION CABLE to this instrument.
The 9461 CONNECTION CABLE is terminated as follows.
See Section 2.2.3 "Connecting the 9641 CONNECTION CABLE (for pulse input)"
(page 29) of Instruction Manual.
7. Connect the 9641 CONNECTION CABLE to the pulse output terminals
of the power meter.
For details of connections, refer to the instructions supplied with the
power meter, and take great care that the connections are correct.
For P1, the white lead is high, and the red lead is low.
8. Turns on the power switch.
See Section 2.3 "Powering On/Off" (page 31) of Instruction Manual.
9. Set the parameters of this instrument.
Shield
(connect to instrument ground)
Ends of the 9641 CONNECTION CABLE
P1 White (H)
P1 Red (L)
P2 Yellow (H)
P2 Black (L)
P3 Green (H)
P3 Blue (L)
P4 Gray (H)
P4 Brown (L)