EasyManua.ls Logo

Hoefer GD2000 - Set the Timer

Hoefer GD2000
30 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
p13
Step 4: Set the timer
The amount of time required for a gel to dry
depends on such factors as gel thickness, gel
concentration, drying temperature, and vacuum
applied. A typical sequencing or 1.5 mm 10%
T gel can be expected to dry in approximately
45 minutes at 80 °C. Larger gels may take 2 to
3 hours. When dry, the thickness of an agarose
gel seen through the silicone flap decreases to
about 1 mm.
1
Press the Time/Temp key to go to the Time mode.
The LED display reads 0:FF and the LED colon blinks
to indicate Elapsed Time.
2
Press either the Up or Down key once to go from
Elapsed Time to Set Time.
3
Press either the Up or Down key to change the
Set Time.
Press Up once to set time for a continuous run. On
a continuous run, the LED display reads r:un and
the heater and vacuum run continuously until you
manually turn them both off.
Press Up again to count up in 15-minute intervals.
Press the Up and Down keys simultaneously to reset
the time to 0. When the Set Time is set to 0, the
display reads 0:FF.
When you have finished setting the time, you
can start the heater and vacuum.
Note:  Agarose gels become brittle
when over-dried.