Open circuit or no output current Blown fuse in protection circuit.
for TE70X or TE77X Replace fuse, (see Accessories in Ordering informa-
tion on page 22).
Smeared or diffuse band patterns If equilibrating before the transfer, shorten or elimi-
nate the equilibration time and/or equilibrate under
coldroom conditions.
If the transfer buffer contains 10% methanol, equili-
brate the gel in transfer buffer for 30 minutes to allow
it to shrink before assembling the stack.
Note: Large proteins may not migrate as readily once
the pore size is slightly reduced.
Take care that the gel does not shift once it contacts
the membrane.
Buffer depletion can change the pH in the transfer
stack, and have a negative effect on the transfer. On
subsequent transfers, either shorten the transfer time,
reduce the current, or increase the number of buffer-
soaked blotting papers in the stack.
Check that the preferred binding surface of the
membrane (if any) contacts the gel.
Uneven band transfer The blotting paper and membrane must be the same size
as the gel or 1 2 mm smaller. Larger sizes will provide an
electrical path for current to bypass the gel solution.
Different proteins will transfer at different rates
depending on size and net charge.