Hospitex Diagnostics EOS Bravo and EOS Bravo Plus Service Manual
39
28) adjust the rotation sense of the sampling needle using the screen of hardware test, pressing
in sequence the keys F5 (service), F3 (diagnostics) and again F3 (hardware test) starting from
the main menu and then evidence the line SN (sampling needle) of the motor’s table using the
arrows up and down.
Press the key F1 (zero motor) to execute a research of the sampling needle which should stop
all in high without the instrument is signing an error (beep). If this is not the case, i.e. the
sampling needle is stops all in down and / or an error is signed, disconnect the connector of
the cabling with the note MSN, exchange between them the two blue coloured cables,
reconnect the connector and repeat this point the verify that the rotation sense has been
effectively changed.
29) adjust the rotation sense of the aspiration needle using the screen of hardware test,
pressing in sequence the keys F5 (service), F3 (diagnostics) and again F3 (hardware test)
starting from the main menu and then evidence the line AN (aspiration needle) of the motor’s
table using the arrows up and down.
Press the key F1 (zero motor) to execute a research of the aspiration needle which should stop
all in high without the instrument is signing an error (beep). If this is not the case, i.e. the
aspiration needle is stops all in down and / or an error is signed, disconnect the connector of
the cabling with the note MAN, exchange between them the two blue coloured cables,
reconnect the connector and repeat this point the verify that the rotation sense has been
effectively changed.