Hospitex Diagnostics EOS Bravo and EOS Bravo Plus Service Manual
40
30) verification of the diluter’s electrovalve functionality using the screen hardware test,
pressing in sequence the keys F5 (service), F3 (diagnostics) and again F3 (hardware test)
starting from main menu and then press the key F4 (valve on /off) and on video we have the
message VALVE ( 1 – 4 ) ?.
Select 1 and verify that the plastic cursor (black and white) present on the central part of the
electrovalve of the diluter is moving correctly. At the end of this operation press another time 1
to leave the cursor in release position.