Hospitex Diagnostics EOS Bravo and EOS Bravo Plus Service Manual
7
MAIN MENU’ > F5 SERVICE > F1 USER PARAMETERS
WASHING STEPS : Peristaltic pump’s steps relative to the first and the second
phase of aspiration, used to wash the cuvette with liquid and air
ASPIRATION STEPS : Peristaltic pump’s steps used to aspirate the remain of the
liquid from the cuvette
OUTER PLATE TEMP. : Parameter to modify the temperature of the plate which holds
the reaction segments
CUVETTE TEMP. : Parameter to modify the temperature of the cuvette
ENABLE REACTION SHAKE : Parameter to enable the function of shake of the plate which
holds the reaction segments
( 0 = NO, 1 = YES )
PRE-HEATING DUTY-CYCLE : Parameter relative to the power impulse relative to the
resistance that is situated around the heating reel
PRINTER MODE : Parameter relative to the printing
0 : 0 = ALL THE PRINTING ARE MADE ON THE INTERNAL
PRINTER
1 : 0 = THE PRINTINGS OF THE SESSION ARE MADE ON
THE INTERNAL PRINTER AND THE PRINTINGS OF THE
REFERTS ARE MADE ON THE EXTERNAL PRINTER
1 : 1 = ALL THE PRINTING ARE MADE ON THE EXTERNAL
PRINTER
AUTOMATIC DATE : Parameter relative to the date and time from the internal PC
0 = NO
1 = ONLY DATE
2 = DATE + HOUR