Hospitex Diagnostics EOS Bravo and EOS Bravo Plus Service Manual
8
MAIN MENU’ > F5 SERVICE > F2 WASHINGS
F1 WASH POSITION : Used to select the number of the hole used for the washings
F2 WASHES : The instrument executes a number of XX washings in automatic
F3 TUBES WASH : The instrument executes a washing cycle of the sampling circuit
F4 CUVETTE WASH : The instrument executes a washing cycle of the aspiration circuit
F5 AUTOZERO : The instrument executes the autozero of all filters on board to the
filter wheel