Performance
Tests
4-36
h.
Repeat
Steps
e
thru
g
until
time
interval
measurements
have
been
taken
for
each
voltage
listed
in
Step
d.
The
repeating
sequence
is:
1.
Set
the
time
interval
to
200
ns.
2.
Increment
the
dc
voltage
by
1
V.
3.
Record
the
new
time
interval
on
the
work
sheet.
i.
Each
value
entered
in
column
‘‘T”’’
is
the
incremental
time
interval
for
a
1
volt
input
step,
plus
200
ns.
The
accumulative
time
interval
is
the
sum
of
all
the
preceeding
incremental
time
intervals,
less
200
ns
per
time
interval.
Find
the
accumulative
time
intervals
(yn)
and
enter
the
results
in
column
‘‘C’’,
and
in
the
‘‘Measured
Results’?
column
on
the
Perfor-
mance
Test
Record.
If,
for
example,
all
the
incremental
time
intervals
are
700
ns,
the
ac-
cumulative
time
intervals
would
be:
0,
500
ns,
1000
ns,
1500
ns,
. .
.
5000
ns
j.
Total
all
the
entries
in
column
‘‘C’’.
k.
Multiply
the
corresponding
entries
in
column
‘‘A’”’
and
column
“‘C’’
together
and
enter
the
results
to
5
decimal
places
in
column
‘‘D’’.
1.
Total
all
the
entries
in
column
‘‘D’’.
m.
The
general
equations
to
find
the
slope
and
y
intercept
for
a
‘‘best
fit
straight
line’
are
rather
involved,
however,
in
this
particular
procedure
they
reduce
to:
ae
DL
eee
3
~
110
=
11
n.
Using
the
values
for
C
and
D
from
the
work
sheet
and
the
formula
from
the
previous
step,
compute
values
for
m
and
b.
o.
Using
the
formula
y
=
mx
+
b,
with
the
specific
values
m
and
b
just
found,
compute
a
value
y
for
each
x
recorded
in
column
‘‘A’’.
Enter
the
results
in
the
‘‘Specification’’
column
on
the
Performance
Test
Record.
p.
Take
the
last
entry
in
the
‘‘Specification”
column
and
multiply
it
by
.005.
Enter
this
number
in
all
the
spaces
in
the
‘‘Tolerance’’
column.
q.
In
order
for
the
3336
to
pass
this
performance
test,
the
‘‘Measured
Results’?
must
equal
the
corresponding
‘‘Specification’’,
plus
or
minus
the
‘‘Tolerance’’.