Loading...
yy
|
@
It
Model
3336A/B/C
The
ac
voltmeter
used
in
this
test
must
be
accurate
to
+.15%
a.
Connect
the
equipment
as
shown
in
Figure
4-2.
Use
the
proper
termination
at
the
voltmeter’s
input.
For
example,
if
you
are
testing
the
600
ohm
output,
use
a
600
ohm
ter-
mination.
NOTE
-hp-
3336
(DUT}
eee
ae
a)
JINN
oo
Jog
ea
ae
Eee
a
eae
|
een
fe
pee
ee
So
Se
i].
AC
VOLIMEIER
ae
ae
Figure
4-2.
Absolute
Amplitude
Accuracy
Equipment
Set
Up.
b.
Refer
to
the
following
table
for
Step
c.
OUTPUT
AMPLITUDE
FREQUENCY
NOMINAL
MINIMUM
MAXIMUM
3336A
752
1500
6090
3336B
75Q
1240
1350
6002
3336C
500
750
c.
Set
up
the
3336
for
each
of
test
conditions
above
and
enter
the
ac
voltmeter
readings
on
the
Performance
Test
Record.
Refer
to
the
minimum
and
maximum
values
above
to
see
7.00
dBm
1.76
dBm
7.00
dBm
7.00
dBm
1.76
dBm
1.76
dBm
7.00
dBm
8.76
dBm
7.00
dBm
10
kHz
50
kHz
10
kHz
10
kHz
50
kHz
50
kHz
10
kHz
10
kHz
10
kHz
if
the
measured
values
meet
the
specifications.
0.6131
0.4743
1.7341
0.6131
0.4312
0.4500
1.7341
0.6130
0.6131
0.6096
0.4716
1.7242
0.6096
0.4288
0.4474
1.7242
0.6095
0.6096
Performance
Tests
0.6166
0.4770
1.7441
0.6166
0.4337
0.4526
1.7441
0.6166
0.6166