FREQuency Subsystem
FREQuency
Subsystem
The
FREQuency
command sets the oscillator frequency. The
FREQuency?
query returns the current test frequency setting.
Command Syntax
FREQuency
2
:CW
3
8
<
:
<
value
>
MIN
MAX
9
=
;
Where,
<
value
>
is the NR1, NR2, or NR3 format
MIN
Sets to the minimum value (20 Hz)
MAX
Sets to the maximum value (1 MHz)
Note
A sux multiplier and a sux unit,
HZ
(hertz), can b e used with this
command. Either
MAHZ
and
MHZ
can b e used as the sux multiplier
for MHz (1E6 Hz).
Example
OUTPUT 717;"FREQ 1KHZ" ! Set to 1 kHz
OUTPUT 717;"FREQ MIN" ! Set to 20 Hz
OUTPUT 717;"FREQ MAX" ! Set to 1 MHz
Query Syntax
FREQuency
2
:CW
3
?
MIN
MAX
Query Response
Returned Format is :
<NR3><NL^END>
Example
10 OUTPUT 717;"FREQ? MIN"
20 ENTER 717;A
30 PRINT A
40 END
Command Reference 8-13