CURRent Subsystem
CURRent Subsystem
The
CURRent
command sets the oscillator's output current level. The
CURRent?
query returns the current oscillator current level.
Command Syntax
CURRent
2
:LEVel
3
8
<
:
<
value
>
MIN
MAX
9
=
;
Where,
<
value
>
is the NR1, NR2 or NR3 format
MIN
Sets or returns the minimum oscillator current level
(50
A)
MAX
Sets or returns the maximum oscillator current level
(When Hi-PW mo de is OFF: 20mA
When Hi-PW mo de is ON :200mA)
In case of ALC ON, this command sets to ALC OFF
and sets the maximum oscillator current level.
Note
A sux multiplier and a sux unit,
A
(amp ere), can be used with
this command. If this query is received when the oscillator level set
to a current level, error
0
230
Data corrupt or stale
will o ccur.
Example
OUTPUT 717;"CURR 10MA" ! Set to 10 mA
OUTPUT 717;"CURR MAX" ! Set to 20 mA
(the HP 4284A standard configuration)
Query Syntax
CURRent
2
:LEVel
3
?
MIN
MAX
Query Response
Returned format is :
<NR3><NL^END>
Example
10 OUTPUT 717;"CURR? MIN"
20 ENTER 717;A
30 PRINT A
40 END
Command Reference 8-15