EasyManua.ls Logo

HP 4284A - Page 258

HP 4284A
462 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OUTPut Subsystem
OUTPut Subsystem
The
OUTPut
subsystem command sets the DC bias monitor function.
Figure 8-4 shows the command tree of the
OUTPut
subsystem group.
Figure 8-4. OUTPut Subsystem Command Tree
:High POWer
The
:High POWer
command sets the high p ower mo de to ON or
OFF, which means that Option 001(Power Amplier/DC Bias) is
valid or invalid when Option 001 is installed. The
:High POWer?
query returns the current high p ower mo de setting. Refer to
Appendix G.
Command Syntax
OUTPut:HPOWer
8
>
>
<
>
>
:
ON
OFF
1
0
9
>
>
=
>
>
;
Where,
1
(decimal 49) When the function is ON
0
(decimal 48) When the function is OFF
Example
OUTPUT 717;"OUTP:HPOW ON"
OUTPUT 717;"OUTP:HPOW 0"
Query Syntax
OUTPut:HPOWer?
Query Response
Returned format is :
<NR1><NL^END>
Example
10 OUTPUT 717;"OUTP:HPOW?"
20 ENTER 717:A
30 PRINT A
40 END
Command Reference 8-17