OUTPut Subsystem
:DC:ISOLation
The
:DC:ISOLation
command sets the bias current isolation to ON
or OFF. The
:DC:ISOLation?
query returns the current setting of
the bias current isolation function. Refer to Appendix G.
Command Syntax
OUTPut:DC:ISOLation
8
>
>
<
>
>
:
ON
OFF
1
0
9
>
>
=
>
>
;
Where,
1
(decimal 49) When the function is ON
0
(decimal 48) When the function is OFF
Example
OUTPUT 717;"OUTP:DC:ISOL ON"
OUTPUT 717;"OUTP:DC:ISOL 0"
Query Syntax
OUTPut:DC:ISOLation?
Query Response
Returned format is :
<NR1><NL^END>
Example
10 OUTPUT 717;"OUTP:DC:ISOL?"
20 ENTER 717:A
30 PRINT A
40 END
8-18 Command Reference