FUNCtion Subsystem
:Source MONitor:IAC
The
:Source MONitor:IAC
command sets the current level monitor
to ON or OFF. The
:Source MONitor:IAC?
query returns the
current ON/OFF condition of the current level monitor.
Command Syntax
FUNCtion:SMONitor:IAC
2
:STATe
3
8
>
>
<
>
>
:
ON
OFF
1
0
9
>
>
=
>
>
;
Where,
1
(decimal 49) When the switchisON
0
(decimal 48) When the switch is OFF
Example
OUTPUT 717;"FUNC:SMON:IAC 1"
Query Syntax
FUNCtion:SMONitor:IAC
2
:STATe
3
?
Query Response
Returned format is :
<NR1><NL^END>
Example
10 OUTPUT 717;"FUNC:SMON:IAC?"
20 ENTER 717;A
30 PRINT A
40 END
8-28 Command Reference