FUNCtion Subsystem
:DEV
<
n
>
:MODE
The
:DEV
<
n
>
:MODE
command sets the deviation measurement
mo de. The
:DEV
<
n
>
:MODE?
query returns the current setting of the
deviation measurement mo de.
Command Syntax
FUNCtion:DEV
<
n
>
:MODE
8
<
:
ABSolute
PERCent
OFF
9
=
;
Where,
ABSolute
1 ABSolute deviation mo de
PERCent
1 % deviation mo de
OFF
Turn the deviation measurement mo de OFF
<
n
>
is :
1
(decimal 49) Deviation mo de setting for primary parameter
2
(decimal 50) Deviation mo de setting for secondary parameter
Example
OUTPUT 717;"FUNC:DEV1:MODE ABS"
OUTPUT 717;"FUNC:DEV2:MODE OFF"
Query Syntax
FUNCtion:DEV
<
n
>
:MODE?
Query Response
Returned format is :
8
<
:
ABS
PERC
OFF
9
=
;
<NL^END>
Example
10 OUTPUT 717;"FUNC:DEV1:MODE?"
20 ENTER 717;A$
30 PRINT A$
40 END