FUNCtion Subsystem
:DEV
<
n
>
:REFerence
The
:DEV
<
n
>
:REFerence
command sets the reference value for
deviation measurement. The
:DEV
<
n
>
:REFerence?
query returns
the current reference values.
Command Syntax
FUNCtion:DEV
<
n
>
:REFerence
<
value
>
Where,
<
value
>
is the NR1, NR2, or NR3 format
<
n
>
is :
1
(decimal 49) Reference value setting for primary parameter
2
(decimal 50) Reference value setting for secondary parameter
Example
OUTPUT 717;"FUNC:DEV1:REF 10"
OUTPUT 717;"FUNC:DEV2:REF 2E-3"
Query Syntax
FUNCtion:DEV
<
n
>
:REFerence?
Query Response
Returned format is :
<NR3><NL^END>
Example
10 OUTPUT 717;"FUNC:DEV1:REF?"
20 ENTER 717;A
30 PRINT A
40 END
:DEV
<
n
>
:REFerence:FILL
The
:DEV
<
n
>
:REFerence:FILL
command executes a single
measurement and enters two measured values (the primary and
secondary parameters) into each of the reference values for the
deviation measurement.
Command Syntax
FUNCtion:DEV
<
n
>
:REFerence:FILL
Where,
<
n
>
is
1
or
2
(Both reference values are measured simultaneously.)
Example
OUTPUT 717;"FUNC:DEV1:REF:FILL"
8-30 Command Reference