INITiate Subsystem
:CONTinuous
The
:CONTinuous
command sets the trigger system to the
CONTinuous ON or OFF condition. In the CONTinuous ON
condition, after reading the measurement data by a controller, the
IDLE STATE is automatically set to the WAIT FOR TRIGGER
STATE. For details refer to \Trigger System" in Chapter 7.
The
:CONTinuous?
query resp onds the current condition of the
CONTinuous ON or OFF.
Command Syntax
INITiate:CONTinuous
8
>
>
<
>
>
:
ON
OFF
1
0
9
>
>
=
>
>
;
Where,
1
(decimal 49) When the function is ON
0
(decimal 48) When the function is OFF
Example
OUTPUT 717;"INIT:CONT ON"
Query Syntax
INITiate:CONTinuous?
Query Response
Returned format is :
<NR1><NL^END>
Example
10 OUTPUT 717;"INIT:CONT?"
20 ENTER 717;A
30 PRINT A
40 END