COrrection Subsystem
:USE:DATA?
The
:USE:DATA?
query returns the OPEN/SHORT/LOAD correction
measurement data at FREQ1, FREQ2 or FREQ3.
Query Syntax
CORRection:USE:DATA?
<
channel number
>
Where,
<
channel number
>
is : 1 to 127 (NR1)
Query Response
Returned format is :
<
open1 A
>
,
<
open1 B
>
,
<
short1 A
>
,
<
short1 B
>
,
<
load1 A
>
,
<
load1 B
>
,
<
open2 A
>
,
<
open2 B
>
,
<
short2 A
>
,
<
short2 B
>
,
<
load2 A
>
,
<
load2 B
>
,
<
open3 A
>
,
<
open3 B
>
,
<
short3 A
>
,
<
short3 B
>
,
<
load3 A
>
,
<
load3 B
>
,
Where,
<
open1/2/3 A
>
NR3 format : primary OPEN correction data at
FREQ1/2/3.
<
open1/2/3 B
>
NR3 format : secondary OPEN correction data at
FREQ1/2/3.
<
short1/2/3 A
>
NR3 format : primary SHORT correction data at
FREQ1/2/3.
<
short1/2/3 B
>
NR3 format : secondary SHORT correction data
at FREQ1/2/3.
<
load1/2/3 A
>
NR3 format : primary LOAD correction data at
FREQ1/2/3.
<
load1/2/3 B
>
NR3 format : secondary LOAD correction data at
FREQ1/2/3.
Example
10 OPTION BASE 1
20 DIM A(18)
30 OUTPUT 717;"CORR:USE:DATA? 89"
40 ENTER 717;A(*)
50 PRINT A(*)
60 END
8-64 Command Reference