G accuracy describ ed in this paragraph applies to the G-B
combination only.
D accuracy
D accuracy
D
e
is given as,
D
e
=
6
A
e
100
Accuracy applies when
D
x
(measured D value)
0.1.
When
D
x
>
0.1, multiply
D
e
by(1+
D
x
).
Q Accuracy
Q accuracy is given as,
6
Q
2
x
2
D
e
1
7
Q
x
2
D
e
Where,
Q
x
is the measured Q v
alue,
D
e
is the relative D accuracy.
Accuracy applies when
Q
x
2
D
e
<
1.
Accuracy
accuracy is given as,
180
2
A
e
100
[
deg
]
G Accuracy
When
D
x
(measured D value)
0.1.
G accuracy is given as,
B
x
2
D
e
[
S
]
B
x
=2
f C
x
=
1
2
f L
x
Where,
B
x
is the measured B value [S],
C
x
is the measured C value [F],
L
x
is the measured L value [H],
D
e
is the relative D accuracy,
f
is the test frequency [Hz].
G accuracy describ ed in this paragraph applies to the C
p
-G and L
p
-G
combinations only.
R
p
Accuracy
When
D
x
(measured D value)
0.1
R
p
accuracy is given as,
General Information 9-9