Table 9-3. Cable Length Factor K
bb
Frequency Cable Length
0m 1m 2m 4m
f
m
100 kHz 1 1+5
2
f
m
1+10
2
f
m
1 +20
2
f
m
100 kHz
<
f
m
300 kHz 1 1+2
2
f
m
1+4
2
f
m
1+8
2
f
m
300 kHz
<
f
m
1 MHz 1 1+0.5
2
f
m
1+1
2
f
m
1+2
2
f
m
f
m
:Test Frequency [MHz]
Table 9-4. Calibration Interpolation Factor K
c
Test Frequency K
c
Direct Calibration Frequencies 0
Other Frequencies 0.0003
Direct Calibration Frequencies are the following 48 frequencies.
Table 9-5. Preset Calibration Frequencies
20 25 30 40 50 60 80 [Hz]
100 120 150 200 250 300 400 500 600 800 [Hz]
1 1.2 1.5 2 2.5 3 4 5 6 8 [kHz]
10 12 15 20 25 30 40 50 60 80 [kHz]
100 120 150 200 250 300 400 500 600 800 [kHz]
1 [MHz]
Table 9-6. Cable Length Factor K
d
Test Signal
level
Cable Length
1m 2m 4m
2V
rms
2.5
2
10
-4
(1+50
2
f
m
)5
2
10
-4
(1+50
2
f
m
)1
2
10
-3
(1+50
2
f
m
)
>
2V
rms
2.5
2
10
-3
(1+16
2
f
m
)5
2
10
-3
(1+16
2
f
m
)1
2
10
-2
(1+16
2
f
m
)
f
m
:Test Frequency [MHz]
Figure 9-6. Temperature Factor K
e
General Information 9-15