Temperature Coefficient
MEDIUM integration time and op erating temp erature at 23
C
6
5
C
Test Signal Level
j
Z
j
,
j
Y
j
,L, C, R D
20 mV
rms
<
0.0025%/
C
<
0.000025/
C
<
20 mV
rms
<
0.0075%/
C
<
0.000075/
C
Settling Time
Frequency (f
m
)
<
70 ms (
f
m
1 kHz)
<
120 ms (100 Hz
f
m
<
1 kHz)
<
160 ms (
f
m
<
100 Hz)
Test Signal Level
<
120 ms
Measurement Range
<
50 ms/range shift (
f
m
1 kHz)
Input Protection
Internal circuit protection, when a charged capacitor is connected to
the UNKNOWN terminals.
The maximum capacitor voltage is:
V
max
=
r
1
C
[
V
]
Where,
V
max
200 V,
C
is in Farads.
Figure 9-7. Maximum Capacitance Voltage
9-22 General Information