Table 10-3.
Test Signal Level/Level-Monitor Test Limits (Hi-PW OFF)
Test Signal
Level
Test Limits
Multimeter Reading Level Monitor Reading
5mV 5mV
6
1.5 mV M.R.
6
(3%+0.5 mV)
10 mV 10 mV
6
2mV M.R.
6
(3%+0.5 mV)
20 mV 20 mV
6
3mV M.R.
6
(3%+0.5 mV)
50 mV 50 mV
6
6mV M.R.
6
(3%+0.5 mV)
100 mV 100 mV
6
11 mV M.R.
6
(3%+0.5 mV)
200 mV 200 mV
6
21 mV M.R.
6
(3%+0.5 mV)
250 mV 250 mV
6
26 mV M.R.
6
(3%+0.5 mV)
500 mV 500 mV
6
51 mV M.R.
6
(3%+0.5 mV)
1V 1V
6
101 mV M.R.
6
(3%+0.5 mV)
2V 2V
6
201 mV M.R.
6
(3%+0.5 mV)
M.R. is the Multimeter Reading for the HP 4284A's output signal
level.
Table 10-4.
Test Signal Level/Level-Monitor Test Limits (Hi-PW ON)
Test Signal
Level
Test Limits
Multimeter Reading Level Monitor Reading
10 mV 10 mV
6
2mV M.R.
6
(3%+0.5 mV)
1V 1V
6
0.11 V M.R.
6
(3%+0.5 mV)
2.5 V 2.5 V
6
0.26 V M.R.
6
(3%+5 mV)
20 V 20 V
6
2.01 V M.R.
6
(3%+5 mV)
M.R. is the Multimeter Reading for the HP 4284A's output signal
level.
10-8 Performance Tests