Impedance
Measurement
Accuracy Test
This test veries the HP 4284A's imp edance measurement accuracy.
Figure 10-6. Impedance Measurement Accuracy Test Setup
Equipment:
10 pF Standard Capacitor HP 16382A
100 pF Standard Capacitor HP 16383A HP 16380A
1000 pF Standard Capacitor HP 16384A
0.01
F Standard Capacitor HP 16385A
0.1
F Standard Capacitor HP 16386A HP 16380C
1
F Standard Capacity HP 16387A
Calibration R-L Standard HP 16074A
Adapter BNC(f ) to BNC(f ) HP PN 1250-0080 4 ea.
Test Leads (1 m) HP 16048A
Test Leads (2 m) HP 16048D (Option 006 only)
Test Leads (4 m) HP 16048E (Option 006 only)
Procedure:
1. Perform a SYSTEM RESET as describ ed on in \System Reset".
2.
Press the
4
MEAS SETUP
5
MENU key and the
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CORRECTION
softkey
to display the
CORRECTION
page.
3. Set the
CORRECTION
page as shown in Figure 10-7.
10-12 Performance Tests