Table 10-7. Impedance Measurement Accuracy Test Limits (2 of 2)
Setting Test Limits (Cp)
Signal
Level
Test
Frequency
0.01
F Standard 0.1
F Standard 1
F Standard
510 mV
1 kHz C.V.
6
0.0082 nF C.V.
6
0.081 nF C.V.
6
0.81 nF
20 mV
1 kHz C.V.
6
0.0188 nF C.V.
6
0.182 nF C.V.
6
1.87 nF
5.1 V
*
1 kHz C.V.
6
0.0132 nF C.V.
6
0.131 nF C.V.
6
1.31 nF
C.V.: Standard's calibration value at 1 kHz.
C.V.
1
: C.V. multiplied by 1.0005
C.V.
2
: C.V. multiplied by 1.0002
*
Option 001 only
15. Set the measurement function to
R-X
.
16. Perform Steps 17 through 19 for all the standards and settings
listed in Table 10-8.
17. Connect the Standard to the HP 4284A's
UNKNOWN
terminals.
18. Press the
4
TRIGGER
5
key.
19. Conrm that the HP 4284A's readings are within the test limits
listed in Table 10-8.
Performance Tests 10-15