Table 10-8.
Impedance Measurement Accuracy Test Limits
Setting Test Limits (R)
Signal
Level
Test
Frequency
100 Standard 1 k Standard
*
510 mV 20 Hz C.V.
6
0.285 C.V.
6
2.81
125 Hz C.V.
6
0.112 C.V.
6
1.11
1 kHz C.V.
6
0.082 C.V.
6
0.81
12.5 kHz C.V.
6
0.132 C.V.
6
1.31
48 kHz C.V.
6
O.132 C.V.
6
1.31
96 kHz C.V.
6
0.132 C.V.
6
1.31
1 MHz C.V.
6
0.154 C.V.
6
1.02
20 mV 20 Hz C.V.
6
0.691 C.V.
6
6.60
1 kHz C.V.
6
O.191 C.V.
6
1.82
1 MHz C.V.
6
0.264 C.V.
6
2.06
5.1 V
**
20 Hz C.V. *0.284 C.V.
6
2.81
1 kHz C.V.
6
O.131 C.V.
6
1.31
1 MHz C.V.
6
0.204 C.V.
6
1.52
C.V.: Standard's calibration value at DC.
*
1 k standard should be measured on the
300
range.
**
Option 001 only
20. Connect the 1 m Test Leads (HP 16048A) to the
UNKNOWN
terminals.
21. Press the
4
MEAS SETUP
5
MENU key and the
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CORRECTION
softkey
to display the
CORRECTION
page.
22. Set the
CABLE
length selection switchto 1 m.
23. Store the
OPEN CORRECTION
data and the
SHORT
CORRECTION
data referring to Steps 4 through 9. In this
pro cedure the OPEN termination and the SHORT termination
should b e connected to the 1 m Test Leads (HP 16048A).
24. Set the HP 4284A to the manual trigger mo de.
25. Set the measurement function to
Cp-D
.
26. Connect the 1000 pF standard to the 1 m Test Leads (HP
16048A)
27. Perform Step 28 through 29 for all the test frequencies listed in
Table 10-9.
10-16 Performance Tests