Single Channel
Correction Mode
When the correction mo de is set to SINGLE, the OPEN/SHORT
interpolation correction data or the FREQ1/FREQ2/FREQ3
correction data are selected automatically, dep ending on the test
frequency and the settings of the OPEN, SHORT, LOAD, FREQ1,
FREQ2 and FREQ3 elds. Table D-1 shows the correction data
selection rules for the SINGLE mo de and corresp onding test
frequencies.
Table D-1. Correction Data Selecting Rule for SINGLE Mode
Test
Frequency
FRQ1/2/3
setting
OPEN SHORT LOAD
ON OFF ON OFF ON OFF
Test Freq=
FRQ1
FRQ1:ON FRQ1
OPEN data
Not
performed
FRQ1
SHORT
data
Not
performed
FREQ1
LOAD data
Not
performed
FRQ1:OFF OPEN data Not
performed
SHORT
data
Not
performed
Not
performed
Not
performed
Test Freq=
FRQ2
FRQ2:ON FRQ2
OPEN data
Not
performed
FRQ2
SHORT
data
Not
performed
FRQ2
LOAD data
Not
performed
FRQ2:OFF OPEN data Not
performed
SHORT
data
Not
performed
Not
performed
Not
performed
Test Freq=
FRQ3
FRQ3:ON FRQ3
OPEN data
Not
performed
FRQ3
SHORT
data
Not
performed
FRQ3
LOAD data
Not
performed
FRQ3:OFF OPEN data Not
performed
SHORT
data
Not
performed
Not
performed
Not
performed
Test Freq
6
=
FRQ1/2/3
FRQ1:ON
FREQ2:ON
FRQ3:ON
OPEN data Not
performed
SHORT
data
Not
performed
Not
performed
Not
performed
FRQ1:OFF
FRQ2:OFF
FRQ3:OFF
OPEN data Not
performed
SHORT
data
Not
performed
Not
performed
Not
performed
Test Freq :Test frequency
OPEN data : OPEN interpolation correction data
SHORT data : SHORTinterp olation correction data
FREQ1/2/3 OPEN data : OPEN correction data at FREQ1/2/3
frequency
FREQ1/2/3 SHORT data : SHORT correction data at FREQ1/2/3
frequency
FREQ1/2/3 LOAD data :LOAD correction data at FREQ1/2/3
frequency
D-2 Correction Data