Multi Channel
Correction Mode
To use the multi channel correction mo de, Option 301 Scanner
Interface must be installed, and the scanner interface function
should b e set to ON from the SYSTEM CONFIG page. When the
correction mo de is set to MULTI, only FREQ1/FREQ2/FREQ3
correction data is used for correction. If the settings of the FREQ1,
FREQ2 and FREQ3 elds are OFF (even the OPEN, SHORT and
LOAD elds are ON), correction will not b e p erformed. Table D-2
shows the correction data selection rules for MULTI mo de and
corresp onding test frequencies.
Table D-2. Correction Data Selecting Rule for MULTI Mode
Test
Frequency
FRQ1/2/3
setting
OPEN SHORT LOAD
ON OFF ON OFF ON OFF
Test Freq=
FRQ1
FRQ1:ON FRQ1
OPEN data
Not
performed
FRQ1
SHORT
data
Not
performed
FRQ1
LOAD data
Not
performed
FRQ1:OFF Not
performed
Not
performed
Not
performed
Not
performed
Not
performed
Not
performed
Test Freq=
FRQ2
FRQ2:ON FRQ2
OPEN data
Not
performed
FRQ2
SHORT
data
Not
performed
FRQ2
LOAD data
Not
performed
FRQ2:OFF Not
performed
Not
performed
Not
performed
Not
performed
Not
performed
Not
performed
Test Freq=
FRQ3
FRQ3:ON FRQ3
OPEN data
Not
performed
FRQ3
SHORT
data
Not
performed
FRQ3
LOAD data
Not
performed
FRQ3:OFF Not
performed
Not
performed
Not
performed
Not
performed
Not
performed
Not
performed
Test Freq
6
=
FRQ1/2/3
FRQ1:ON
FREQ2:ON
FRQ3:ON
Not
performed
Not
performed
Not
performed
Not
performed
Not
performed
Not
performed
FRQ1:OFF
FRQ2:OFF
FRQ3:OFF
Not
performed
Not
performed
Not
performed
Not
performed
Not
performed
Not
performed
Test Freq :Test frequency
OPEN data : OPEN interpolation correction data
SHORT data : SHORTinterp olation correction data
FREQ1/2/3 OPEN data : OPEN correction data at FREQ1/2/3
frequency
FREQ1/2/3 SHORT data : SHORT correction data at FREQ1/2/3
frequency
FREQ1/2/3 LOAD data :LOAD correction data at FREQ1/2/3
frequency
Correction Data D-3