CORRECTION
FUNCTION SETTING
To set the correction function or to obtain the correction data, use
the following summary of the correction function setting elds with
their HP-IB commands (given in the short form). (A) through (F)
shown in the gure corresp ond to the description of each eld.
Field Softkey Command Description
(A) OPEN ON CORR:OPEN:STATON :Performs OPEN correction using OPEN
data or FREQ1/2/3 OPEN data.
OFF CORR:OPEN:STAT OFF : Not p erform OPEN correction.
MEAS
OPEN
CORR:OPEN : Obtains OPEN interpolation data
(B) SHORT ON CORR:SHOR:STATON :Performs SHORT correction using
SHORT data or FREQ1/2/3 SHORT data.
OFF CORR:SHOR:STAT OFF : Not p erform SHORT correction.
MEAS CORR:SHOR : Obtains SHORTinterpolation SHORT
data.
(C) LOAD ON CORR:LOAD:STATON :Performs LOAD correction using
FREQ1/2/3 LOAD data.
OFF CORR:LOAD:STAT OFF : Not p erform LOAD correction.
D-4 Correction Data