• 
• 
•• 
Model 436A 
Performance Tests 
PERFORMANCE TESTS 
4-10. 
ZERO CARRYOVER TEST 
(cont'd) 
8.  Wait for 
the 
Power Meter's  digital readout 
to 
stabilize and verify 
that 
the 
indication 
observed is within 
the 
limits shown on the table below.  Then set 
the 
POWER 
Meter RANGE 
HOLD 
switch 
to 
off 
(out). 
9.  Repeat steps 
6, 
7, and 8 with 
the 
Range Calibrator RANGE switch set, 
in 
tum, 
to 
1mW,10 
mW, 
and 
100 
mW. 
Verify 
that 
the Power Meter autoranges properly, 
and 
that 
the 
indication observed 
on 
each range 
is 
within 
the 
limits shown in 
Table 4-1. 
Table 
4-1. 
Zero 
Carryover 
Autorange 
Digital 
Readout 
Results 
Range 
Calibrator 
Results 
and 
Power Meter 
Range 
Min 
Actual 
Max 
lOOµW 
-0.2 
0.2 
lmW 
-.002 
.002 
lOmW 
-0.02 
0.02 
lOOmW 
-00.2 
00.2 
4-11. 
INSTRUMENT 
ACCURACY 
TEST 
SPECIFICATION:  WATT 
MODE: 
dBm 
MODE: 
dB (REL) 
MODE: 
±0.5% 
in Ranges 1 through 
5. 
±0.02 
dB 
±0.001 dB/°C 
in Ranges 1 through 5. 
±0.02 
dB 
±0.001 dB/°C 
in Ranges 1 through 5. 
NOTE 
The dB (REL) specifications 
are 
for within-range measurements. 
For 
range-to-range accuracy, add the uncertainty associated with 
the range in which the reference 
was 
entered, 
to 
the uncertainty 
associated with 
the 
range in which the measurement 
was 
made. 
For example, 
if 
a reference 
is 
entered in Range 1 and a 
measure-
ment 
is 
made 
in 
Range 
5, 
the total uncertainty 
is 
±0.04 
(Range 1 
±0.02 
+Range 5± 
0.02 
= 
± 
0_04). 
" 
DESCRIPTION: 
After 
the 
Power Meter 
is 
initially 
cal'ibrated 
on the 1 
mW 
range, the digital readout 
is 
monitored as the Range Calibrator 
is 
adj~ed 
to 
provide reference inputs correspond-
ing 
to 
each 
of 
the Power 
Meter operating 
ran'ges~, 
-~ 
'-, 
4-3