5-6
ADJUSTMENTS
10.
11.
12.
13.
14.
Press
the
PRESET/LOCAL
key,
then
the
ENTER
key.
Press
the
ZERO
key,
and
allow
time
(5-15
seconds)
for
the
zeroing
routine
to
finish.
Set
the
range
calibrator’s
FUNCTION
switch
to
CALIBRATE.
Press
the
Power
Meter’s
CAL
(SHIFTed
ZERO)
key.
Using
(<),
(+),
GD,
or
CO),
modify
the
REF
CF
to
read
100.0%
Press
the
ENTER
key.
Set
the
range
calibrator’s
RANGE
to
10
mW.
Adjust
A4RI11
(RNG
4
SHP)
for
a
reading
of
10.00
+0.0!1
mW
on
the
Power
Meter
display.
Set
the
range
calibrator’s
RANGE
to
100
mW.
Adjust
A4R112
(RNG
5
SHP)
for
a
reading
of
100.0
+0.1
mW
on
the
Power
Meter
display.
Repeat
steps
10
through
13
to
check
that
interaction
between
steps
has
not
caused
a
shift
in
settings.