EasyManua.ls Logo

HP 437B - Page 14

HP 437B
57 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HP
437B
Procedure
5-8
ADJUSTMENTS
Connect
the
equipment
as
shown
in
Figure
5-3.
Set
up
the
counter
to
measure
frequency.
Set
the
Power
Meter
LINE
switch
to
ON.
Press
the
PRESET/LOCAL
key,
then
the
ENTER
key.
Press
the
PWR
REF
(SHIFTed
(J)
key.
Observe
the
indication
on
the
frequency
counter.
If
it
is
50.0
+0.5
MHz
no
adjustment
of
the
power
reference
oscillator
frequency
is
necessary.
If
it is
not
within
these
limits,
adjust
the
power
reference
oscillator
frequency
as
described
in
steps
6
and
7.
y
Remove
the
Power
Meter
bottom
cover.
Adjust
A4L5
(FREQ)
to
obtain
a
50.00
+0.5
MHz
indication
on
the
frequency
counter.

Related product manuals