HP
437B
Procedure
5-8
ADJUSTMENTS
Connect
the
equipment
as
shown
in
Figure
5-3.
Set
up
the
counter
to
measure
frequency.
Set
the
Power
Meter
LINE
switch
to
ON.
Press
the
PRESET/LOCAL
key,
then
the
ENTER
key.
Press
the
PWR
REF
(SHIFTed
(J)
key.
Observe
the
indication
on
the
frequency
counter.
If
it
is
50.0
+0.5
MHz
no
adjustment
of
the
power
reference
oscillator
frequency
is
necessary.
If
it is
not
within
these
limits,
adjust
the
power
reference
oscillator
frequency
as
described
in
steps
6
and
7.
y
Remove
the
Power
Meter
bottom
cover.
Adjust
A4L5
(FREQ)
to
obtain
a
50.00
+0.5
MHz
indication
on
the
frequency
counter.