HP 437B - ANALOG BOARD TROUBLESHOOTING (continued)

HP 437B
57 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HP
437B
Service
Sheet
BD1
MICROPROCESSOR
J
CONTROL
Figure
8-3.
A4U16,
U18
Pinout
Diagram
The
digital
control
signals
for
the
Analog
board
have
now
been
verified.
Continue
with
the
troubleshooting
procedure
for
Service
Sheet
A4a.
Service
Sheet
A4a
troubleshooting
1.
Press
the
Power
Meter’s
SET
RANGE
key.
Select
range
3.
2.
Press
the
dBm/W
key
for
a
reading
in
watts.
3.
Set
the
range
calibrator
as
follows:
FUNCTION
ceiscswiiieevecieianeeden
sess
CALIBRATE
RANGE.
osei
adie
ces
pad
aadudenoeta
wee
cohesaee
1
mW
4.
Verify
that
the
level
at
A4TP10
(C-GND)
is
equal
to
A4TP9
(A-GND)
+5.00
mV.
5.
Set
the
oscilloscope
as
follows:
Vertical
Sensitivity
.........
ccc
eee
ce
cece
eee
eeeeenes
5V/div
SWECD
i
idavs
oa.
cewie
Saeed
Se
ercie
weeded
Oo40
Wass
Hee
Madre
1
ms/div
COU
PLN
iceceaet
spire
Ree
Oa
os ea
tees
Ras
OHNE
ae
Nioa
le
oe
DC
TriQQering.
2
..ccscsseseesvaws
vee
eieaedaneal
eee
yess
Internal
6.
Connect
the
oscilloscope
to
A4TP3
(220
Hz).
Verify
the
signal
shown
in
Figure
8-4.
SERVICE
§-33