EasyManua.ls Logo

HP 5245L - Page 45

HP 5245L
270 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model
5245L
Table
3-2. In-Cabinet
Performance
Check
(cont'd)
3. Print
Command:
Negative
step
from +13
volts
to 0
volts, DC-coupled.
a.
Connect
Oscilloscope
to DIGITAL
RECORDER
jack
(Jl1)
pin
48.
b.
Set Counter
controls
as
follows:
SAMPLE
RATE
slightly clockwise
out of POWER
OFF
SENSITIVITY
to
CHECK
TIME
BASE
to 10
ms
FI]NCTION
to
FREQUENCY
c.
The
Oscilloscope
should display the
print
command
step
(-13
volts for
each counting
cycle).
Record
the amplitude
of this negative
step on test card.
4.
Hold-Off
Requirements:
+15V
min., +25Y
mrx.
from
chassis
ground
(tOOO-ofrm
source).
a.
Set Counter
controls
as
follows:
SAMPLE
RATE
slightly
clockwise
out of POWER
OFF
SENSITTVITY
to
CHECK
TIME
BASE
to 10
s
STORAGE
(on
rear
panel)
to non-storage
FUNCTION
to FREQUENCY
b. With
DC Power
Supply,
apply hold-off voltage, +25
volts
in
series with a
1000-ohm resistor
to
DIGITAL
RECORDER,jack
(J11)
Pin22
on the rear
panel.
The
Counter
should
stop
until
the
hold-off
voltage
is
removed from
Jtt,
pin
22.
Record hold-off
voltage on test
card.
c.
Repeat
step b, using +15
hold-off voltage.
Record hold-off
voltage on test
card.
Note
this
check
can
be made
using
any hold-off voltage from +15
volts
to
+25
volts.
GATE INDICATOR.
Front
panel
indication
of main
gate
'ropen"
state
a.
Set Counter
controls
as follows:
SENSITIVITY
to
CHECK
SAMPLE
RATE
slightly clockwise out of POWER
OFF
TIME
BASE
to
10 s
FIINCTION
to MANUAL
STOP
-
LEVEL
+
to
PRESET
b.
Set Counter
FUNCTION
switch to
MANUAL
START, the Counter
gate
lamp
should turn
on.
Record
results
on
test card.
RE-SET CAPABILITY.
A momentary
control on the front
panel
that returns
both
the displayed and
internal
count
to zero.
a.
Set Counter
controls
as follows:
SENSITTVITY
to CHECK
SAMPLE
RATE
slightly
clockwise
out of POWER
OFF
TIME
BASE TO
10s
FI]NCTION
to
FREQUENCY
-
LEVEL
+
to PRESET
b. Push front panel
RESET
pushbutton.
Counter
should reset
and then
start
a new
count.
Record
result
on
test card.
Section
Itr
Table
3-2
02349-1
3-
11

Other manuals for HP 5245L