EasyManua.ls Logo

HP 6296A - Page 40

HP 6296A
61 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table
5-4.
Low
Output
Voltage
Troubleshooting
(Continued)
Step
Measure
Response
Probable
Cause
s
Voltage
between
+s
and
11
a.
Less
positive
than
a.
QlOOB
open
+0.9V
QlOOA
shorted
Q802
or
Q80l
shoned
b,
+O.
9V
to
+l.
SV
b.
Proceed
to
Step
6
6
Voltage
between
+s
and
21
a.
OV
or
positive
a.
Q303
shorted
0302
shorted
R3
l
O
shorted
Table
S-5.
Preregulator/Control
Circuit
Troubleshooting
Step
Measure
Response
Probable
Cause
l
Waveform
between
4
and
3
of
a.
Normal
firing
pulse
a.
CRS02-CRS04
defective
T700
R50l-RS02
open
CRSOl, CR503, T800
defective
b.
No
or
abnormal
fir-
b.
T700
open
ing
pulse
CRSOO
shorted
Proceed
to
Step
2
2
Waveform
between
90
and
47
a.
Zero
or
small
posi-
a.
Q7 00
shorted
tive
voltage
C703
shorted
Q701
open
R703,
primary
T700
open
Proceed
to
Step
3
b.
+16
to
+20
Volt
level
b.
Q700
open
R703
shorted
Q70l
shoned
Proceed
to
Step
3
c.
Waveform
distorted
c.
Proceed
to
Step
3
3
Waveform
between
90
and
59
a.
Amplitude
incorrect
a.
Q702
defective
R707,R708,
R713
incor-
rect
value
or
open
C700,
CR710,
C711
defec-
tive
b.
Period
incorrect
b.
CR709
defective
Proceed
to
Step
4
4
Waveform
between
90
and
52
a.
Amplitude
incorrect
a.
CR708, CR709, R702
de-
fective
b.
Period
incorrect
b.
CR700
through
CR703
defective
s
Waveform
between
90
and
54
a.
Amplitude
incorrect
a.
R700, R701,
C701
defec-
tive
b.
Period
incorrect
b
CR704
throuqh
CR707
de-
fective
5-9

Related product manuals