IACER Srl 32 MNPG53-07
Guidance and manufacturer's declaration ELECTROMAGNETIC IMMUNITY
FOR ALL DEVICES AND SYSTEMS
MAG2000 is intended for use in the electromagnetic environment specified
below. The user or operator of MAG2000 should assure that is used in such
environment.
Immunity test
Test level
IEC 60601
Compliance
level
Electromagnetic
environment
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6kV; +8kV at
contact
±8kV; +15kV in
air
±6kV; ±8kV
at contact
±8kV; +15kV
in air
Floors should be wood,
concrete or ceramic
tile. If floor is covered
with synthetic material,
the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2kV
for power
supply lines
±2kV
for power
supply lines
Mains power quality
should be at least that
one of a typical
commercial or hospital
environment.
Impulses
IEC 61000-4-5
±1kV
line - line
±1kV
line - line
Mains power quality
should be at least that
one of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dips of
U
T
)
for 0,5 cycle
<5% U
T
(>95% dips of
U
T
)
for 1 cycle
70% U
T
(30% dips of
U
T
)
for 25 cycles
<5% U
T
(>95% dips of
U
T
)
<5% U
T
(>95% dips of
U
T
)
for 0,5 cycle
<5% U
T
(>95% dips of
U
T
)
for 1 cycle
70% U
T
(30% dips of
U
T
)
for 25 cycles
<5% U
T
(>95% dips of
U
T
)
Mains power quality
should be at least that
one of a typical
commercial or hospital
environment.
If the user of the
MAG2000 requires
continued operating
during power mains
interruptions, it is
recommended that
MAG2000 be powered
from an uninterruptible
power supply (UPS) or a
battery