MKS Baratron
®
Type
622B/623B/626B
Absolute Pressure Transducers
(505)872-0037
idealvac.com
idealvac.com

Table of Contents

Questions and Answers

  • M
    Manuel ThorntonAug 11, 2025
    What to do if Ideal Vacuum Transducer shows overrange signal?
    • P
      Pamela GomezAug 11, 2025
      If your Ideal Vacuum Transducer shows an overrange positive or negative signal, it may be due to a shorted transducer or a damaged cable connecting the transducer to the electronics module. Check the supply voltages at the terminal block. Carefully inspect the cable and transducer for any signs of damage. If damage is found, replace the cable or transducer.
  • S
    Stephanie EwingAug 21, 2025
    Why Ideal Vacuum Transducer has unstable zero output?
    • C
      Courtney BauerAug 21, 2025
      If you're experiencing an unstable zero output on your Ideal Vacuum Transducer, it could be due to ambient temperature issues. The ambient temperature might be too high or varying too much. Make sure the ambient temperature is within the product's specified requirements, as detailed in Appendix A: Product Specifications, page 49.
  • S
    Stephen RobertsAug 31, 2025
    What to do if Ideal Vacuum Transducer measurement slowly goes positive?
    • D
      dsmithAug 31, 2025
      If the measurement from your Ideal Vacuum Transducer slowly goes positive over time, this could be caused by overpressure or a build-up of contamination in the P x cavity. In this case, you should return the transducer to MKS for servicing or replacement.