16 17
/// Pipetting
Pipetting recommendations
Observe “Technical data“�
The pipette, pipette tip and liquid to be pipetted must have reached the ambient temperature�
Only aspirate liquid if a pipette tip is attached to the tip cone�
Do not touch the pipette tip at the tip orifice�
Change the pipette tip before aspirating different liquids�
Pre-rinse the pipette tip several times before use�
Hold the pipette upright when aspirating liquid�
Dip the pipette tip only a few millimeters into the liquid�
Move the pipetting button slowly and smoothly
After aspiration, wait 1 - 3 seconds to remove the pipette tip from the liquid�
Drops outside the pipette tip must be wiped off the vessel wall�
Place the pipette tip steeply against the vessel wall while dispensing the liquid and wipe the
pipette tip off the vessel wall�
Do not hold the pipette in your hand when you do not need it to avoid the transfer of hand
heat�
Always hold the pipette upright to prevent liquid from entering the pipette body and causing
corrosion�
Turning the multifunction button outside the specified volume range may destroy the display
mechanism and damage the pipette�