If „I external“ is enabled, it is not possible to carry out leakage current
measurement by means of Test socket 1.
„Timer RPE 10A“
Duration of 10 A voltage drop and 10 A earth bond measurement can be set:
• Press the MENU key; the main menu is opened.
• Use ▲ and ▼ keys to select submenu „Set-up“ and press ENTER.
• Use ▲ and ▼ keys to select item „Timer RPE 10A“ and press ENTER.
• Use DEL key to erase current setting.
• Use 0 ÷ 9 keys to enter required new setting (1 ÷ 15 s; if you set value to 0, the
measurement will be running continually until you either disconnect test leads or
internal circuitry overheats). You can use the DEL key to erase incorrectly entered
characters.
• Confirm with the ENTER key.
„Erase memory “
This item erases all measurement results stored in main instrument’s memory! Data in
the buffer are not erased.
Press the ENTER key and release it. Following menu is displayed: „Enter=Yes Esc=No“.
Confirm with the ENTER key. Abandon erasing procedure with the ESC key.
„Nominal voltage“
You can set insulation measurement voltage to 500V, 250V, 100V with the ENTER key.
„PE cross-section“
Timer RPE 10A
15 sec
PE cross-section
1.0 mm2
Erase memory
Erase memory?
Enter=Yes Esc=No
500 V